Product reference

YD9800 Eddy Current Displacement Sensor

Probe and conditioner selection, measurement ranges and environmental specifications.

On this page

Catalog specifications; confirm the ordered configuration and applicable explosion-protection documentation.

Technical specifications

Parameter Specification
Resolution 0.05% FS
Frequency response 0–10 kHz
Linearity error ≤ 1% FSO
Temperature drift ≤ 0.06%/°C
Operating temperature −50 to +175°C
Amplitude response −1% at 1 kHz; −5% at 10 kHz
Supply, negative voltage output −18 to −24 VDC
Supply, current or positive voltage output +18 to +30 VDC
Consumption ≤ 12 mA, excluding output current
Explosion-protection marking Ex ia IIB T6 Ga
Protection IP67

Probe measurement ranges

Probe diameter (mm) Linear range (mm) Linearity error Minimum target dimension (mm; terminology unconfirmed)
φ5 1 ≤ ±1% φ15
φ8 2 ≤ ±1% φ25
φ11 4 ≤ ±1% φ35
φ18 8 ≤ ±1.5% φ45
φ25 12 ≤ ±1.5% φ50
φ50 25 ≤ ±2% φ100

The catalog's general linearity figure and the diameter-specific errors are separate specifications. Use the error for the selected probe; do not apply the single general figure to every range. The original heading translated here as minimum target dimension has unclear terminology; confirm its meaning before specifying a target surface.

Probe ordering options

YD9800-A□□-B□□-C□□-D□□-E□□-F□□-G□□

Parameter Specification
A — Probe diameter 01: φ5; 02: φ8; 03: φ11; 04: φ18; 05: φ25; 06: φ50
B — Thread 01: M8×1; 02: M10×1; 03: M14×1.5; 04: M30×2; 05: custom
C — Mounting 01: standard; 02: reverse; 03: side cable exit; 04: flange
D — Unthreaded length 00: 00 mm; 01: 01 mm; 02: 02 mm; …; 10: 10 mm
E — Overall housing length 20: 20 mm; 50: 50 mm; …; 250: 250 mm
F — Probe cable + extension cable 01: 3 + 0 m; 02: 1 + 4 m; 03: 3 + 6 m
G — Armor 00: armored; 01: unarmored

Only the printed codes are listed. An ellipsis in the catalog does not establish the omitted intermediate ordering codes.

Conditioner ordering options

CWY-DO-A□□-B□□-C□□

Parameter Specification
A — Range 01: 0–1 mm; 02: 0–2 mm; 03: 0–4 mm; 04: 0–8 mm; 05: 0–12 mm; 06: 0–25 mm
B — System cable length 03: 3 m system; 05: 5 m system; 09: 9 m system
C — Output 01: 4–20 mA; 02: −2 to −18 V; 03: 0–5 V; 04: other output

Only the printed codes are listed. An ellipsis in the catalog does not establish the omitted intermediate ordering codes.

Selection drawing and limitations

The selection drawing identifies the probe, probe cable, extension cable and armor by the ordering-table letters. It does not specify additional universal fixed dimensions. The conditioner designation is printed as CWY-DO; whether the final character is the letter O or the digit zero remains unconfirmed. The YD9800 drawing also identifies the wrench flats and separate probe-cable and extension-cable portions of F.