Product reference

YD66 Portable Wireless Inspection Instrument

Wireless vibration, non-contact temperature and reflective laser speed measurement for machinery inspection.

On this page

Catalog specifications; confirm the ordered variant and applicable explosion-protection documentation.

Inspection functions and applications

YD66 combines Bluetooth data transmission with vibration inspection, non-contact infrared temperature measurement and laser speed measurement. The catalog describes simultaneous acquisition of acceleration, velocity and displacement overall values, plus the waveform for one selected measurement type. Magnetic mounting allows the instrument to be moved between inspection locations.

Applications include condition assessment of industrial rotating machinery in power generation, chemical processing, cement and water services. Cable-free vibration measurement avoids sensor cable and plug wear. Wireless charging avoids repeated insertion of a charging connector. Low-power sleep/wake behavior is advertised, although the original wording of that feature is unclear.

Hardware and environment

Parameter Specification
Battery Built-in 3.7 V, 600 mAh lithium battery
Continuous operation 20 hours after a full charge
Communication Bluetooth 4.2
Mounting Magnetic
Protection type Standard: none; explosion-protected: Exib IIC T5 Gb
Protection rating, as printed P66; confirmation required
Materials Stainless steel; plastic
Dimensions, length × width × height 33 × 40 × 90 mm
Weight 150 g
Reliable transmission distance Within 5 m
Antenna Internal
Operating temperature −20 to +70°C

Vibration measurement

Measurement Range and accuracy Frequency response
Acceleration 0–200 m/s² peak, ±5% 10 Hz–2 kHz
Velocity 0–200 mm/s RMS, ±5% 10 Hz–1 kHz
Displacement 0–2000 μm peak-to-peak, ±5% 10 Hz–1 kHz

Sampling frequency: 5120 Hz; 12800 Hz optional.

Infrared temperature measurement

Parameter Specification
Method Non-contact infrared
Range −32 to +420°C
Distance-to-spot ratio D:S = 25:1
Response time < 500 ms
Accuracy, as printed ±1%°C or 1%, whichever is greater; expression unconfirmed
Adjustable parameter 0.1–1.0; the catalog label translates as emission, but its exact parameter name is unconfirmed

The temperature diagram labels distance D in millimeters as 0, 400, 600, 800 and 1000, and target diameter S as 13, 20, 30, 35 and 40 respectively. The relationship between these examples and the stated 25:1 ratio is unconfirmed; do not derive further measurement limits from the diagram.

Laser speed measurement

Parameter Specification
Method Reflected laser light; apply a reflective patch to the shaft before measurement
Speed range 50–8000 r/min
Accuracy ±(0.02%n + 1) r/min
Effective measuring distance Within 50 cm

The component drawing identifies the infrared lens, laser emitter, laser receiver and magnetic base, and shows a wireless connection to a handheld terminal. The speed illustration identifies the rotating part, reflective patch and reflective photoelectric sensor.

Specification limitations

The rating P66 is reproduced without adding an I. The temperature-accuracy expression and the 0.1–1.0 parameter label remain unresolved; they have not been replaced with assumed conventional specifications. The overview gives Exib IIC T5, while the specification table gives Exib IIC T5 Gb; confirm the marking and documents for the selected variant.